The
Ghajar Guide® One Piece
(#01-0001)
THE GHAJAR GUIDE® One Piece consists of an integrated tripod and lumen (Figure 2) This product is available in French 10 (Large, #O1-0001). Instructions for Use
2. Hold the catheter loaded guide (see "Loading the Catheter") firmly against the skull or scalp above the burr hole with one hand. 3. With the free hand, manually advance the catheter into the ventricle. The intracerebral catheter length will vary for different size craniums. In the normal adult the optimal length is 5cm . 4. Grasp the catheter through the base of the guide with shod forceps to prevent inadvertent movement. 5. Holding the catheter in place lift the guide and stylet off in one motion. Test for cerebrospinal fluid manometrics.
Match the catheter to the insert to insure proper fit. Place the appropriate insert, with the flange oriented as in Figure 1, into the tripod base before loading the catheter Nonflanged Catheter (Figure 4) 1. With stylet in place, insert catheter through proximal lumen of THE GHAJAR GUIDE.
1. Remove stylet from catheter (Figure 5a) 2. Insert the nonflanged end of the catheter through lumen of the guide from the base (Figure 5b) 3. Place stylet in catheter (Figure 5c)
Indications THE GHAJAR GUIDE® is designed to be used as a device for anterior placement of an intraventricular catheter at ninety degrees to the skull surface.
Warnings This product is for SINGLE USE ONLY: DO NOT RESTERILIZE. Use aseptic technique in a phases of handling. Neurodynamics will not be responsible for any product that is resterilized, nor will we accept for credit or exchange any product that has been opened but not used.
Sterility As long as the individual package of THE GHAJAR GUIDE® is not opened or damaged the product is sterile.
Caution Federal Law (U.S.A. ) restricts the sale of this device to a physician or by his order.
Warranty Neurodynamics, Inc., warrants that this medical device is free from defects in both materials and workmanship. THE ABOVE WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES, EITHER EXPRESS OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Suitability for use of the medical device for any surgical procedure shall be determined by the user. Neurodynamics shall not be liable for incidental or consequential damages of any kind.
Copyright ©1999-2003. All Rights Reserved. May not be duplicated or distributed in any form. Designed and Powered by MIS, Inc.
|